The device was returned to stryker sustainability solutions for evaluation.Upon visual inspection of the received complaint device, evidence of clinical use was identified.The distal end of the blade was fractured and separated from the rest of the device.Inspection of the broken blade coating revealed signs of erosion near the blade fracture and the distal end of the blade as a result of activating against solid/hard materials.The teflon pad, jaw, handle and contact rings appear to be intact.A review of the dhr supports that the device met all inspection and test criteria prior to release from stryker.Therefore, the most likely root causes are: jaws/blade subassembly damage, incidental and prolonged activation against solid surfaces, such as bone or plastic.The instructions for use (ifu) state: use the torque wrench (already mounted to the shaft) to tighten the blade onto the hand piece.Turn the torque wrench clockwise while holding only the gray hand piece until it clicks twice indicating that sufficient torque has been applied to secure the blade.Do not attempt to bend, sharpen, or otherwise alter the shape of the blade.Doing so may cause blade failure and user or patient injury.Avoid contact with any and all other instruments while the instrument is activated.Contact with staples, clips or other instruments while the instrument is activated may result in cracked or broken blades.Note: do not use any other means than the torque wrench to attach or detach the instrument from the hand piece.Note: do not torque the instrument by hand without the torque wrench or damage may occur to the hand piece.The reported event will continue to be monitored through post-market surveillance.
|
|
It was reported that the harmonic scalpel blade broke off inside the patient.It was retrieved laparoscopically, no additional procedure needed.There was no patient injury, medical intervention, or extended procedure time reported.These are commonly used devices that are readily available.
|