Model Number 504112BL
Medical Device Problem Codes Failure to Osseointegrate (1863); Osseointegration Problem (3003)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 09/10/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A good oral hygiene is recommended for a successful and long-lasting osseointegration.An evaluation of the implant by the manufacturer has shown no indications of a product defect or unknown risks.
 
Event or Problem Description
Inadequate bone qualiy/bone quantity and poor oral hygiene were reported.Bone quality at the time of implant failure type ii.Time of loss in relation to the implantation was before prosthetic restoration.Primary stability was achieved.Osseointegration was not achieved.Augmetation procedure was not performed at the time of surgery.Patient is a smoker.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFINITY OCTAGON IMPLANT
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
HAGER & MEISINGER GMBH
hansemannstrasse 10
neuss nordrhein-westfalen, germany D-414 68
GM  D-41468
MDR Report Key9232361
Report Number1287163-2019-00674
Device Sequence Number14074380
Product Code DZE
UDI-Device IdentifierE0HM504112BL1
UDI-PublicE0HM504112BL1
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K143539
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2019
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 10/21/2019
Report Date (Section F) 09/30/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number504112BL
Device Catalogue Number504112BL
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/30/2019
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date09/30/2019
Event Location Other
Date Report to Manufacturer09/30/2019
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/30/2019
Initial Report FDA Received Date10/24/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age58 YR
-
-