MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARD® TOUCHLESS® PLUS UNISEX PRE-LUBRICATED RUBBER URETHRAL CATHETER KIT; COUDE TOUCHLESS CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARD® TOUCHLESS® PLUS UNISEX PRE-LUBRICATED RUBBER URETHRAL CATHETER KIT; COUDE TOUCHLESS CATHETER Back to Search Results
Model Number 4A7044
Medical Device Problem Code Incorrect Measurement (1383)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event or Problem Description
It was reported that the catheter appeared shorter than normal causing a difficult insertion process.The patient was unable to use the catheter.No medical intervention was required.
 
Additional Manufacturer Narrative
The reported event was confirmed.A bard touchless plus kit was returned opened, containing its insert, two gloves, waterproof underpad, latex intermittent catheter within the 1100cc collection bag, and unopened providine-iodine swabsticks.The catheter contained white markings.The catheter was removed from the bag and measured.The catheter length was14.75+0.125+0.125/4 = 14.90625" and was determined to be below specification per procedure.A potential root cause for this failure could be "catheter is too short, not able to be inserted far enough".The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "instructions for use: preparing for catheterization: 1.Open kit and remove contents.2.The gloves are not necessary to maintain aseptic conditions during the procedure.The gloves are for the operator's protection.Male/female catheterization: 1.Remove cap from insertion tip while squeezing tabs.Save cap for closing the bag.2.Slide the catheter halfway into the insertion tip.3.Male - prepare the male urethral meatus and the surrounding area with povidone-iodine swabs provided.Female - prepare the female urethral meatus by holding the outer labia apart and prep the urethral meatus and surrounding area with povidone iodine swabs provided.4.Male - holding the penis, advance the insertion tip into the urethra no further than the flange base.Female - spread the inner labia, advance the insertion tip into the urethra no further than the flange base.Release the inner labia.5.Place your nondominant hand on the finger control guide to stabilize catheter in urethra.With your dominant hand, grasp catheter through bag approximately 1" below the finger control guide and push catheter into urethra.The catheter should be introduced by short, repetitive pushing motions.Repeat motions until catheter reaches bladder and urine starts to flow.6.Allow urine to flow freely, making certain the catheter guide is elevated at least 4" above lower portion of the bag.Allow flow until bladder is empty or until bag is filled.Precaution: it is recommended that the collection bag be held.The catheter could possibly separate from the collection bag when urine increases weight of the bag.7.Withdraw the catheter from the urethra.Remove the remaining portion of the catheter from the collection bag by pulling it through the insertion tip.Note: the catheter is designed to pass through the insertion tip.8.The filled collection bag may be closed by replacing the cap over the insertion tip.Make sure the cap snaps on securely.Specimen collection: to collect a specimen, obtain a sample cup/tube and alcohol wipe.Remove guide tip by pulling tab(s) upward from bag.Wipe port with alcohol.Pour specimen through port at top of bag into cup/tube.Draining bag: remove insertion tip by pulling tab(s) upward from bag.Drain urine through port at top of bag.".
 
Event or Problem Description
It was reported that the catheter appeared shorter than normal causing a difficult insertion process.The patient was unable to use the catheter.No medical intervention was required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® TOUCHLESS® PLUS UNISEX PRE-LUBRICATED RUBBER URETHRAL CATHETER KIT
Common Device Name
COUDE TOUCHLESS CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key9234135
Report Number1018233-2019-06763
Device Sequence Number11571043
Product Code FCM
UDI-Device Identifier00801741030536
UDI-Public(01)00801741030536
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
K910653
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer,other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 12/06/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date11/30/2021
Device Model Number4A7044
Device Catalogue Number4A7044
Device Lot NumberNGDQ2849
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/16/2019
Initial Date Received by Manufacturer 10/04/2019
Supplement Date Received by Manufacturer12/03/2019
Initial Report FDA Received Date10/24/2019
Supplement Report FDA Received Date12/06/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-