• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION NC EMERGE; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION NC EMERGE; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number 7214
Medical Device Problem Code Entrapment of Device (1212)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/27/2019
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that catheter entrapment occurred.A 3.00mm x 20mm nc emerge balloon catheter was used in a procedure.However, the balloon got stuck on the wire.The physician was able to pull everything out.No patient complications were reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: returned product consisted of an nc emerge balloon catheter.The shaft, hypotube, tip and balloon were microscopically and visually examined.There were numerous kinks.There was contrast in the inflation lumen and balloon.The balloon was loosely folded.Microscopic inspection revealed tip damage.The device was soaked in a water bath for 18 days to loosen the blood and contrast in the device.The device was prepped with an inflation device filled with water and connected to the inflation port to inflate the device.No issues detected.The device was functionally tested with a.014" guidewire.No issues detected.Inspection of the remainder of the device presented no other damage or irregularities.Product analysis did not confirm the reported difficulties with the guidewire.
 
Event or Problem Description
It was reported that catheter entrapment occurred.A 3.00mm x 20mm nc emerge balloon catheter was used in a procedure.However, the balloon got stuck on the wire.The physician was able to pull everything out.No patient complications were reported.It was further reported that: the multivessel percutaneous coronary intervention (pci) was performed with multiple stenosis to the left anterior descending (lad) artery, first diagonal branch (d1) and second obtuse marginal branch (om2).A pci was performed on the 80% stenosed, 20mm long, target lesion between the proximal lad to mid lad.A non-bsc wire passed through the lesion, angioplasty was performed using a 2.25mm x 15mm nc emerge balloon catheter at 14 atmospheres for 15 seconds.However, the balloon ruptured and got separated.The physician deployed the 2.75mm x 32mm synergy drug-eluting stent to the affected area.Post-stent angioplasty was performed using a 3.00mm x 15mm nc emerge at 18 atmospheres for 10 seconds, but the balloon got stuck on the wire.Another non-bsc balloon was used for angioplasty.The stent became thrombosed and was treated with angioplasty and a 2.5mm x 28mm synergy drug-eluting stent deployed to overlap the first placed stent.Angioplasty was performed using a non-bsc balloon.There was a 0% residual stenosis post intervention.A pci was performed on the 90% stenosed target lesion located in the ostial first diagonal to the first diagonal branch (d1).A non-bsc wire passed through the lesion.Angioplasty was performed using a non-bsc balloon.A 2.25mm x 15mm nc emerge balloon catheter was advanced for dilatation at 18 atmospheres for 10 seconds.A 2.25mm x 38mm synergy drug-eluting stent drug-eluting stent was successfully placed.Post-stent angioplasty was performed using a non-bsc balloon.The stent became thrombosed intraprocedure.Angioplasty was performed with a 2.0 x 12mm emerge balloon at 10 atmospheres for 12 seconds and a non-bsc balloon.The stent thrombosis was treated with balloon angioplasty.There was 0% residual stenosis post intervention.A pci was performed on the 95% stenosed target lesion located in the ostial om2 to om2 lesion.A non-bsc wire passed through lesion.The pre-interventional distal flow was normal (timi 3).Angioplasty were performed using a 2.25 x 15mm nc emerge balloon catheter at 10atmospheres for 15 seconds and two non-bsc balloons.The shaft of the 2.25 nc emerge severed and was left in the patient's om2.Then a non-bsc stent was placed successfully.The post-interventional distal flow was decreased.There was a 99% residual stenosis post intervention.The multilevel pci to the lad, d1 and om2 was complicated by the balloon issues and lad stent thrombosis.Myocardial infarction, cardiogenic shock, and cardiac arrest were encountered.The cardiac arrest was treated with cardiopulmonary resuscitation (cpr), heart pump and extracorporeal membrane oxygenation (ecmo) and the patient was sent to the coronary care unit.Successful return of spontaneous circulation with ongoing ecmo support.There was pulmonary hemorrhage secondary to anticoagulation and cpr.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: returned product consisted of an nc emerge balloon catheter.The shaft, hypotube, tip and balloon were microscopically and visually examined.There were numerous kinks.There was contrast in the inflation lumen and balloon.The balloon was loosely folded.Microscopic inspection revealed tip damage.The device was soaked in a water bath for 18 days to loosen the blood and contrast in the device.The device was prepped with an inflation device filled with water and connected to the inflation port to inflate the device.No issues detected.The device was functionally tested with a.014" guidewire.No issues detected.Inspection of the remainder of the device presented no other damage or irregularities.Product analysis did not confirm the reported difficulties with the guidewire.
 
Event or Problem Description
It was reported that catheter entrapment occurred.A 3.00mm x 20mm nc emerge balloon catheter was used in a procedure.However, the balloon got stuck on the wire.The physician was able to pull everything out.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC EMERGE
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key9235233
Report Number2134265-2019-12711
Device Sequence Number2181179
Product Code LOX
UDI-Device Identifier08714729846659
UDI-Public08714729846659
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K141236
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/17/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number7214
Device Catalogue Number7214
Device Lot Number0024006380
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/01/2019
Initial Date Received by Manufacturer 10/08/2019
Supplement Date Received by Manufacturer12/09/2019
11/25/2019
Initial Report FDA Received Date10/24/2019
Supplement Report FDA Received Date12/10/2019
12/17/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age52 YR
-
-