ANIMAS LLC ANIMAS VIBE; INSULIN INFUSION PUMP Back to Search Results
Device Problems Electrical /Electronic Property Problem (1198); Moisture Damage (1405)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.Animas has conducted a review of the device history record for this pump and confirmed that it was operating within required specifications at the time of release.
 
Event Description
On (b)(6) 2019, the reporter contacted animas, alleging a power (moisture ingress) issue.It was reported that the pump had no power and there was moisture behind the display.There was no indication that the product caused or contributed to an adverse event.This complaint is being reported because the user may be unaware that the pump has lost power, leading to under delivery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANIMAS VIBE
Type of Device
INSULIN INFUSION PUMP
Manufacturer (Section D)
ANIMAS LLC
965 chesterbrook blvd
wayne PA 19087
Manufacturer (Section G)
ANIMAS LLC
965 chesterbrook blvd
wayne PA 19087
Manufacturer Contact
kristen lopolito
965 chesterbrook blvd
wayne, PA 19087
MDR Report Key9236217
MDR Text Key170324516
Report Number2531779-2019-05855
Device Sequence Number1
Product Code OYC
Combination Product (y/n)Y
Reporter Country CodeCA
PMA/PMN Number
P130007
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,foreign
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 09/29/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Initial Date Manufacturer Received 09/29/2019
Initial Date FDA Received10/24/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age34 YR
-
-