BIOSENSE WEBSTER INC THERMOCOOL® SMART TOUCH® SF BI-DIRECTIONAL NAVIGATION CATHETER; CARDIAC ABLATION PERCUTANEOUS CATHETER Back to Search Results
Catalog Number D134803
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Complete Heart Block (2627); No Code Available (3191)
Date of Event 10/02/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The biosense webster inc.Product analysis lab received the device for evaluation on 10/15/2019.The analysis has begun but is not completed at this time.When the investigational analysis has been completed, a supplemental 3500a report will be submitted.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's ref.No: (b)(4).
 
Event or Problem Description
It was reported that a male patient underwent an atrial fibrillation (afib) procedure with a thermocool® smart touch® sf bi-directional navigation catheter and suffered a heart block av which required a permanent implantable pacemaker.During the procedure, a junctional escape rhythm was observed and a heart block was discovered.The temporary pacemaker was set.The patient was reported to be in stable condition and a 24 hour hold had been put on to see if the node would reinitiate activity.The permanent pacemaker was to be set if no activity was seen during the 24 hour hold.The permanent pacemaker was implanted the next day.The patient required extended hospital stay and permanent damage was sustained by the av node requiring a permanent implantable pacemaker.The adverse event was discovered during use of biosense webster products.The physician¿s opinion on the cause of this adverse event is that it was related to procedure while ablating a flutter in the left atrium near the left atrial septum.
 
Additional Manufacturer Narrative
Investigation summary: it was reported that a male patient underwent an atrial fibrillation (afib) procedure with a thermocool® smart touch® sf bi-directional navigation catheter.During the procedure, a junctional escape rhythm was observed and a heart block was discovered.The temporary pacemaker was set.The patient was reported to be in stable condition and a 24 hour hold had been put on to see if the node would reinitiate activity.The permanent pacemaker was to be set if no activity was seen during the 24 hour hold.The permanent pacemaker was implanted the next day.The patient required extended hospital stay and permanent damage was sustained by the av node requiring a permanent implantable pacemaker.The adverse event was discovered during use of biosense webster products.The device was visually inspected and it was found in good conditions.The magnetic sensor was tested on carto and the catheter was properly visualized and no errors were observed.Then, the force sensor was tested and it was working properly, the force values were observed within specifications.Then, electrical test was performed on the catheter and it was found within specifications.No electrical malfunction was observed.Additionally, the catheter was tested on the generator and the temperature and impedance values were observed within specifications.Then, the irrigation and deflection test was performed and it was found within specifications, the catheter was irrigating and deflecting correctly.A manufacturing record evaluation was performed and no internal actions related to the reported complaint were identified.The catheter passed all specifications.The root cause of the adverse event remains unknown.The instructions for use (ifu) states that careful catheter manipulation must be performed to avoid cardiac damage, perforation or tamponade.Manufacturer's reference number: (b)(4).
 
Additional Manufacturer Narrative
During an internal review on december 3, 2019, a correction was noted on the patient codes.It was assessed to add the patient code of ¿surgical intervention¿ therefore, ¿no code available¿ was added to : patient code.Patient code of ¿no code available¿ represents ¿surgical intervention." manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THERMOCOOL® SMART TOUCH® SF BI-DIRECTIONAL NAVIGATION CATHETER
Common Device Name
CARDIAC ABLATION PERCUTANEOUS CATHETER
Manufacturer (Section D)
BIOSENSE WEBSTER INC
33 technology drive
irvine CA 92618
MDR Report Key9236778
Report Number2029046-2019-03814
Device Sequence Number19316613
Product Code LPB
UDI-Device Identifier10846835010169
UDI-Public10846835010169
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P030031
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/02/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/12/2020
Device Catalogue NumberD134803
Device Lot Number30220578L
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/15/2019
Initial Date Received by Manufacturer 10/02/2019
Supplement Date Received by Manufacturer11/04/2019
12/03/2019
Initial Report FDA Received Date10/25/2019
Supplement Report FDA Received Date11/27/2019
12/03/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
6FR CS,F,10 POLE,12 PIN, AUTOID; 71CM TRANSSEPTAL NEEDLE; LASSO NAV 2515,12P SPLITHANDLE; SOUNDSTAR ECO SMS 8F CATHETER
Outcome Attributed to Adverse Event Hospitalization; Required Intervention; Disability;
-
-