• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDACTA INTERNATIONAL SA STEM: QUADRA-H CEMENTLESS, HA COATED STD STEM SIZE 2, SHORT NECK; CEMENTLESS STEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDACTA INTERNATIONAL SA STEM: QUADRA-H CEMENTLESS, HA COATED STD STEM SIZE 2, SHORT NECK; CEMENTLESS STEM Back to Search Results
Catalog Number 01.12.22SN
Health Effect - Clinical Code Inadequate Osseointegration (2646)
Date of Event 09/27/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Batch review performed on 03 october 2019.Lot 181779: (b)(4) items manufactured and released on 18-jul-2018.Expiration date: 2023-07-03.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been already sold without any other similar reported event.Clinical evaluation performed by medical affairs manager: hip revision surgery performed 8 months after cementless total hip arthroplasty in a (b)(6) woman.No information concerning patient general health status, comorbidities and level of activity is available.Radiographic image provided shows the presence of radiolucent lines in gruen zones 1 and 7.Aseptic loosening is a possible literature described adverse event after cementless total hip arthroplasty and causes are often unknown.The reason of this event cannot be determined.
 
Event or Problem Description
Revision following the loosening of the stem 8 months after primary.The surgery was completed successfully.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEM: QUADRA-H CEMENTLESS, HA COATED STD STEM SIZE 2, SHORT NECK
Common Device Name
CEMENTLESS STEM
Manufacturer (Section D)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, 6874
SZ  6874
MDR Report Key9237117
Report Number3005180920-2019-00903
Device Sequence Number11154271
Product Code JDI
UDI-Device Identifier07630030802621
UDI-Public07630030802621
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K082792
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 10/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/03/2023
Device Catalogue Number01.12.22SN
Device Lot Number181779
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/27/2019
Initial Report FDA Received Date10/25/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age74 YR
-
-