Catalog Number 151820029
Medical Device Problem Code Loss of or Failure to Bond (1068)
Health Effect - Clinical Code Pain (1994)
Date of Event 12/18/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Medical records received 17 june 2019 and were reviewed 02 october 2019 for mdr reportability.On (b)(6) 2017, the patient underwent total right knee arthroplasty due to osteoarthritis.The patella was resurfaced.The surgeon reported no intraoperative complications.The patient was implanted with the depuy knee system and unknown bone cement.On (b)(6) 2018, the patient underwent a right knee revision due to loosening and pain.The surgeon indicated the tibial component was loose and about to come out.He reported the patellar button came off very easily and was revised.There were no concerns with the femoral component, yet it was revised.The patient was implanted with a competitor system and there were no complications to the procedure.Doi: (b)(6) 2017.Dor: (b)(6) 2018, (rt knee).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provision of 21 cfr, part 803.The report may be based on the information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Additional narrative: product complaint # (b)(4).Investigation summary: the device associated with this report was not received for examination.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot: null.Device history batch: null.Device history review: null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE MEDIAL DOME PAT 29MM
Common Device Name
ATTUNE IMPLANT : KNEE PATELLA
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
MDR Report Key9238264
Report Number1818910-2019-112026
Device Sequence Number16555552
Product Code JWH
UDI-Device Identifier10603295056669
UDI-Public10603295056669
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K103756
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2018
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/02/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/31/2017
Device Catalogue Number151820029
Device Lot Number3565459
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 10/02/2019
Supplement Date Received by Manufacturer11/25/2019
Initial Report FDA Received Date10/25/2019
Supplement Report FDA Received Date11/26/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ATTUNE CR FB INSRT SZ 4 5MM.; ATTUNE CR FEM RT SZ 4 NAR CEM.; ATTUNE CR FB INSRT SZ 4 5MM; ATTUNE CR FEM RT SZ 4 NAR CEM
Outcome Attributed to Adverse Event Required Intervention;
Patient Age65 YR
-
-