Model Number SAFB-SM/INTL
Medical Device Problem Code Low Readings (2460)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/09/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during use, the device¿s measurement value started to dropped at the middle of the surgery and remained to the lowest value.No increased in value was seen even during patient recovery.Sensor was replaced with a new one and the value increased.There was no patient injury.
 
Additional Manufacturer Narrative
The returned sensor was reported as ¿during use, the device¿s measurement value started to dropped at the middle of the surgery and remained to the lowest value.No increased in value was seen even during patient recovery.Sensor was replaced with a new one and the value increased.There was no patient injury¿.The test-bed device was connected to ac power.The ac power illuminated.The test-bed device was powered on and passed.The test-bed device was tested for proper operation using a pre-amplifier and a field test device and was observed to operate as specified.A visual inspection was carried out on the returned sensor.No anomalies were observed.The sensor was connected to a reusable sensor cable, the pre-amplifier and the test-bed device.The sensor failed with the following error message 'poor signal quality' coming up on the display screen, and no values could be obtained.The sensor was tested as per the post assembly test procedure, where it was observed that the sensor was passing ¿noise¿ but failing ¿shallow¿ at ¿functional¿.The failing component was located on the adult sensor.Which by design, was a flexible printed circuit board.The sensor was sent to the vendor for analysis.The device failed to meet specification as it was received or made available for evaluation.The investigation isolated the failure of the observed condition to the defective sensor.Information has been added to the database and trends will continue to be monitored.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INVOS
Common Device Name
OXIMETER, TISSUE SATURATION
Manufacturer (Section D)
NELLCOR PURITAN BENNETT IRELAND
micheal collins rd mervue
galway 20023
EI  20023
Manufacturer (Section G)
NELLCOR PURITAN BENNETT IRELAND
micheal collins rd mervue
galway 20023
EI   20023
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view,mn, CO 55112
7635265677
MDR Report Key9238686
Report Number2936999-2019-00893
Device Sequence Number16187373
Product Code MUD
UDI-Device Identifier20884521104225
UDI-Public20884521104225
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K082327
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSAFB-SM/INTL
Device Catalogue NumberSAFB-SM/INTL
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/07/2020
Is the Reporter a Health Professional? Yes
Initial Date Received by Manufacturer 10/09/2019
Supplement Date Received by Manufacturer01/29/2020
08/08/2024
Initial Report FDA Received Date10/25/2019
Supplement Report FDA Received Date01/31/2020
08/14/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-