• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET 3I OSSEOTITE® TAPERED CERTAIN® IMPLANT 6 X 8.5MM; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET 3I OSSEOTITE® TAPERED CERTAIN® IMPLANT 6 X 8.5MM; DENTAL IMPLANT Back to Search Results
Model Number INT685
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Information (3190)
Date of Event 09/17/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Zimmer biomet complaint (b)(4).A summary investigation has been completed for lack of primary stability events that do not allege a deficiency with the implant and identified that the reported event is likely due to biological factors which have an adverse effect on implant stability or is related to surgical technique and insertion torque.Due to a wide range of external factors (non-design/non-manufacturing related), identifying a definitive root cause is generally not possible.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
 
Event or Problem Description
Doctor reported that they were unable to achieve primary stability for implant int685 in tooth site #14.Site was bone grafted and patient sent home to heal.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OSSEOTITE® TAPERED CERTAIN® IMPLANT 6 X 8.5MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
MDR Report Key9238988
Report Number0001038806-2019-01249
Device Sequence Number2409163
Product Code DZE
UDI-Device Identifier00844868006787
UDI-Public(01)00844868006787(17)211128(10)2016110531(241)INT685
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K063341
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2019
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 10/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/28/2021
Device Model NumberINT685
Device Catalogue NumberINT685
Device Lot Number2016110531
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/25/2019
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/25/2019
Initial Report FDA Received Date10/25/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age62 YR
-
-