Additional Manufacturer Narrative
|
|
Initial reporter - unknown contact.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Device report from synthes reports an event in (b)(6) as follows: it was reported that on (b)(6) 2019, during a knee surgery when the needle of the omnispan meniscal repair, 0deg was inserted into the joint it made a double deployment, as per the surgeon the needles were removed and they used a new omnispan meniscal repair, 0deg but the same issue occurred, a double deployment was made.No patient consequence and no surgical delay was reported.No additional information was provided.This report is for one (1) omnispan meniscal repair, 0deg.This is report 1 of 2 for (b)(4).
|
| |
|
Event or Problem Description
|
|
Updated: during the implant the correct order of triggers was pulled.The needle was in the correct orientation.The needle did not over penetrate the meniscus.The surgeon maintained the appropriate technique and position.No implants remained in the patient.The procedure was completed using a new omnispan implant.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Investigation summary: the complaint device is not being returned, it was discarded by the customer, therefore unavailable for a physical evaluation.This complaint cannot be confirmed.A manufacturing record evaluation was performed for the finished device l356652 number, and no non-conformance's were identified.Since the complaint device was discarded, we cannot determine a root cause for the reported failure.If additional information is received in the future, we will reopen the complaint and perform the investigation as appropriate.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Device history: the batch review l356652 showed that the (b)(4) devices were conformed to in process and finished goods specifications when released to stock on april 18th 2017.Expiration date: march 31st 2020 (according to retained labels).No nc was opened.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Corrected data: after additional review, it was found that this impacted product and complaint is a duplicate of another complaint; (b)(4).Please see mfr: 1221934-2019-59223.
|
| |
|