• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE Back to Search Results
Model Number V60
Device Problem Failure of Device to Self-Test (2937)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6) 2019.Date of report: 10/25/2019.The remote technician performed troubleshooting with the customer.The customer reported that when the flow was set to 0, they received a reading of -21.The call was routed to the field.The manufacturer's field service engineer (fse) performed troubleshooting and confirmed the reported issue.The fse also noted that the unit declared a low leak co2 rebreathing alarm which was duplicated during troubleshooting.The fse replaced the flow sensor assembly and both issues were resolved.The unit passed performance verification testing.No parts were returned to failure investigation; therefore, the root cause of the reported issue could not be determined.
 
Event Description
It was reported that the customer was unable to perform leak testing on the unit.There was no patient involvement.
 
Manufacturer Narrative
Date received by manufacturer: 29dec2019.Date of this report: (b)(6) 2019.The gas delivery system (gds) was returned for failure analysis.A visual inspection showed no signs of damage or contamination.During analysis, it was determined that the air flow sensor failure was caused by the u1 component drifting out of specification.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V60 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, INC
2271 cosmos court
carlsbad CA 92011
MDR Report Key9240321
MDR Text Key170699429
Report Number2031642-2019-10259
Device Sequence Number1
Product Code MNT
Combination Product (y/n)N
PMA/PMN Number
K082660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 10/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/25/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberV60
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/19/2019
Was the Report Sent to FDA? No
Date Manufacturer Received10/16/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-