Catalog Number 275010140
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The device is available for evaluation.Investigation will be conducted.Follow up will be filed with the investigation results.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Libertas: it was reported that on (b)(6) 2019, the 6mm dual-lead tap was found to be broken during reverse logistics audit of returned device at millstone.There was no patient involvement and no additional information is available.This complaint involves one (1) device.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Visual inspection of the returned device found the ball tip feeler shaft was bent.The ball tip was broken, and not returned to the manufacturer.No other issues were identified with the returned components of the device.A cursory inspection of the complaint file for the device that was found damaged does not point to a manufacturing issue.Therefore a device history record is not required.A definitive assignable root cause for the post manufacturing damage could not be determined based on the provided information.This complaint is confirmed however no product design issues or manufacturing discrepancies that would contribute to the reported complaint condition were identified during this investigation.No new, unique or different patient harms were identified because of this evaluation.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BALL TIP FEELER STRAIGHT
Common Device Name
PROBE
Manufacturer (Section D)
DEPUY SPINE INC
325 paramount drive
raynham MA 02767
MDR Report Key9240601
Report Number1526439-2019-52295
Device Sequence Number1646213
Product Code HXB
UDI-Device Identifier10705034194066
UDI-Public10705034194066
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/30/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number275010140
Device Lot Number0110V
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/03/2019
Initial Date Received by Manufacturer 09/30/2019
Supplement Date Received by Manufacturer11/05/2019
Initial Report FDA Received Date10/25/2019
Supplement Report FDA Received Date11/07/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-