A review of the device history record has been completed.No deviations or non-conformances noted.Device evaluation: analysis of the returned device identified: opening curved radius leak test and microscopic analysis was performed which identified: striated punctured on plug strap disk shell, striated opening on radius, creases fold, wear abrasion and observed void >1 mm in non-textured, valve-to shell-bond area.The fill test inspection was performed, the result is no blockage.Based on the device analysis the final assessment is: a striated opening on radius due to surgical damage consist in the use of some surgical tool.A striated punctured on plug strap disk shell due to surgical damage like consist in the use of some surgical tool.Reason for reoperation: deflation.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.This is a known potential adverse event addressed in the product labeling.
|