• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DAVIS & GECK CARIBE LTD CHROMIC GUT; SUTURE, ABSORBABLE, NATURAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVIS & GECK CARIBE LTD CHROMIC GUT; SUTURE, ABSORBABLE, NATURAL Back to Search Results
Model Number UG-202
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/12/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, when making the knots during a circumcision procedure, the suture was released even when there were no resistance to from the knot.Another suture was used to complete the case.There was no patient injury.
 
Additional Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) received thirty-one devices.A tactile and microscopic inspection of the sutures was performed with no abnormalities.A needle attachment test was performed on the sealed samples, and the results met the specifications for this product.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CHROMIC GUT
Common Device Name
SUTURE, ABSORBABLE, NATURAL
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DO  0101
MDR Report Key9241138
Report Number9612501-2019-02074
Device Sequence Number150753
Product Code GAL
UDI-Device Identifier10884521056138
UDI-Public10884521056138
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K885018
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2021
Device Model NumberUG-202
Device Catalogue NumberUG-202
Device Lot NumberD8A1798ZX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/05/2020
Initial Date Received by Manufacturer 10/10/2019
Supplement Date Received by Manufacturer05/13/2020
Initial Report FDA Received Date10/25/2019
Supplement Report FDA Received Date05/29/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-