Model Number MODEL 3000
Medical Device Problem Code Computer Software Problem (1112)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/02/2019
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a patient's had a diagnostics test result in high impedance.The patient was referred for a full revision due to the high impedance.It was determined that the cause of this false high impedance message was a software error on m3000 (b)(4) software; when system diagnostics were performed by the user with m3000 (b)(4) software on m102/102r generators (normal mode output current >0 ma), normal mode diagnostics would actually be performed instead.The execution of normal diagnostics instead of system diagnostics is not apparent to the user, and the returned dcdc code was being compared against system diagnostic thresholds.No known surgical intervention occurred to date.No additional information has been received to date.
 
Event or Problem Description
It was reported that the generator was prophylactically replaced.The explanted generator has not been received to date.No additional information has been received to date.
 
Event or Problem Description
It was reported that the patient was interrogated with version 1.0 software with normal impedance.No additional information has been received to date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Common Device Name
PROGRAMMING SOFTWARE
Manufacturer (Section D)
LIVANOVA USA, INC.
100 cyberonics blvd
houston TX 77058
MDR Report Key9241388
Report Number1644487-2019-02068
Device Sequence Number11174131
Product Code LYJ
UDI-Device Identifier05425025750429
UDI-Public05425025750429
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
P970003
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Remedial Action Notification
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMODEL 3000
Was Device Available for Evaluation? No
Event Location Other
Initial Date Received by Manufacturer 10/02/2019
Supplement Date Received by Manufacturer10/29/2019
12/12/2019
Initial Report FDA Received Date10/25/2019
Supplement Report FDA Received Date11/21/2019
01/06/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age26 YR
-
-