| Model Number |
TN,DXH 800 HEMATOLOGY SYSTEM |
| Medical Device Problem Code |
Output Problem (3005)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
09/25/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Evaluation codes - method, result, and conclusion codes updated.The field service engineer (fse) inspected the instrument, did not identify any specific failure, and was unable to replicate the reported issue.Per review of the data provided by technical support on 31-oct-2019, all latron, start up, and control data were within specifications.Technical support recommended the instrument have the latron nrbc al2 whole blood calibration procedure done because the latron calibration values were within specification but were on the high side.The fse went onsite to perform nrbc and latron calibration on 12-nov-2019.He performed the recommend calibration successfully and provided the calibration data.Review of the data could not confirm an instrument malfunction.A cause for the erroneous unflagged erroneous could not be confirmed.The erroneous un-flagged erroneous results could not be confirmed by the fse.Evidence provided does not allow to confirm if erroneous results were generated or identify a conclusive failure mode.Bec internal identifier - (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Raw data and additional information regarding the results and patient outcome has been requested.This is an ongoing investigation and an update will be provided when new information becomes available.Bec internal identifier: case (b)(4).
|
| |
|
Event or Problem Description
|
|
The customer reported that multiple erroneous high platelet (plt) results were generated by their unicel dxh 800 coulter cellular analysis instrument.The customer stated that there were both flagged an unflagged results.Additional information has been requested, with the current information available it is not known how many patients were affected by this event.The customer reported that some results had been released by the time she became aware of this issue.Some of the patients had their results corrected.There was no report of impact to patient or change in patient treatment connected to this event.
|
| |
|
Search Alerts/Recalls
|