Correction to evaluation codes method, result and conclusion.Device evaluation summary: one pressure solenoid printed circuit board (psol pcb) was returned to medtronic for further analysis.A visual inspection was carried out, no anomalies were observed.The psol pcb was assembled into the failure investigation test-bed device for analysis and the unit was powered on.The unit passed the power on self test and calibrations successfully without any errors.The unit was placed into ventilation mode for 24 hours, no errors/alarms were observed.Performed and passed functional testing on the psol pcb.Conclusion: no fault found.The investigation found the device to function normally.The event will be included in trending and monitoring.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|