| Model Number |
FGS-0450 |
| Medical Device Problem Code |
Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Information (3190)
|
| Date of Event |
10/03/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, the recorder failed to switch on and was not recognized by the computer.The recorder worked correctly on the previous procedure.A repeat procedure will be performed.There was no user or patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: this report is based on information provided by medtronic investigation personnel.The bravo was received for evaluation.The returned sample did not meet specification as received by medtronic.The visual inspection found the cover of the on/off switch stuck inside.The customer reported the recorder the recorder failed to switch on and was not recognized by the computer.The reported condition was confirmed.The investigation found the switch on/off of the recorder stuck inside, it is soft and crooked, but still functional.This led to the reported problem.The investigation identified the cause of the reported event to be power button.Corrective actions have been implemented to mitigate this issue.If information is provided in the future, a supplemental report will be issued.
|
| |
|