MAUDE Adverse Event Report: MERIT MEDICAL SYSTEMS, INC. IMPRESS PERIPHERAL CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MERIT MEDICAL SYSTEMS, INC. IMPRESS PERIPHERAL CATHETER Back to Search Results
Catalog Number 41103510PIG-NB
Medical Device Problem Codes Break (1069); Fracture (1260); Material Separation (1562)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/01/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The suspect device has been returned for evaluation.The product was examined visually.The complaint is confirmed.No definitive root cause could be determined however, it is likely that significant force was applied to the device during use.A search of the complaint database found no similar complaints for this lot number.The device history record was reviewed and no exception documents were found.
 
Additional Manufacturer Narrative
The suspect device has been returned for evaluation.A follow up will be submitted when the evaluation is complete.
 
Event or Problem Description
The account alleges that during a vascular procedure, the catheter tip detached within the patient.The physician had acquired arterial access and had successfully crossed the targeted arterial lesion.During catheter manipulations over a guidewire the tip detached.The physician was able to successfully capture and remove the foreign body with a vascular snare device.No additional patient consequences to report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPRESS PERIPHERAL CATHETER
Common Device Name
PERIPHERAL CATHETER
Manufacturer (Section D)
MERIT MEDICAL SYSTEMS, INC.
14646 kirby drive
houston TX 77047
MDR Report Key9242432
Report Number3010665433-2019-00070
Device Sequence Number1892100
Product Code DQO
UDI-Device Identifier00884450022436
UDI-Public00884450022436
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K053171
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2021
Device Catalogue Number41103510PIG-NB
Device Lot NumberE1490033
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/21/2019
Initial Date Received by Manufacturer 10/03/2019
Supplement Date Received by Manufacturer11/11/2019
Initial Report FDA Received Date10/25/2019
Supplement Report FDA Received Date11/14/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-