| Model Number |
MMT-522LNAH |
| Medical Device Problem Code |
Failure to Cycle (1142)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/23/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer's mother reported via phone call that the customer's insulin pump had compromised force sensor system.Customer¿s blood glucose level was 167 mg/dl.Customer stated that they were unable to exit the preparing to prime loop and the rewind message occurred after an alarm or alert and they were stuck in rewind complete or bolus delivery loop.Customer stated that they did not receive 2nd series of beeps nor did numbers appeared on the screen.The customer was also advised that the insulin pump needed to be replaced and was advised to discontinue use of the insulin pump and to revert to the backup plan.The device will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device stuck in loop a33 error due to loose or protruded drive support disk noted.Unable to perform any test.
|
| |
|