• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. ETHLN BLK 18IN 4-0 S/A P-3 PRM MP; SUTURE, NONABSORBABLE, SYNTHETIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. ETHLN BLK 18IN 4-0 S/A P-3 PRM MP; SUTURE, NONABSORBABLE, SYNTHETIC Back to Search Results
Catalog Number 699G
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.The product upon which this medwatch is based has been received, however, the product evaluation is not yet complete.Any further information derived from the evaluation will be submitted in a supplemental 3500a form.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.When event occurred- pre-op- before use on patient? or intra-op-during use on patient while suturing? please explain/clarify what do you mean by " suture coming apart while the physician was trying to use them"? ----did the suture pull off/detach/separate from the needle? or did the suture thread break into 2 pieces? please confirm specific number of devices (total qty actually involved with reported issue) for each code/lot reported that failed for this event? note: events reported in 2210968-2019-89096, 2210968-2019-89098, 2210968-2019-89099, and 2210968-2019-89100.
 
Event or Problem Description
It was reported that a patient underwent an unknown procedure on an unknown date and suture was used.During the procedure, the suture came apart while the physician was trying to use it.There were no adverse patient consequences reported.Additional information has been requested.
 
Additional Manufacturer Narrative
(b)(4).One empty opened box, two bottom overwrap pieces and nine unopened samples of product code 699, lot pdb848 were received for analysis.The complaint sample was not received for evaluation.During the visual inspection of samples, no defects were found on the packages.The samples were opened, and the swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along of the strands and no defects were observed.A functional test was performed and the pull force were above the minimum requirements.The manufacturing records were reviewed, and the manufacturing/packaging criteria were met prior to the release of this batch.Per the condition of the samples, no pull off suture needle was found, and the tested samples met the finished goods requirements.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ETHLN BLK 18IN 4-0 S/A P-3 PRM MP
Common Device Name
SUTURE, NONABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
p.o. box 151, route 22 west
somerville NJ 08876
MDR Report Key9243781
Report Number2210968-2019-89097
Device Sequence Number3763829
Product Code GAR
UDI-Device Identifier10705031008625
UDI-Public10705031008625
Combination Product (Y/N)N
Initial Reporter StateNH
Initial Reporter CountryUS
PMA/510(K) Number
K946173
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/30/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2024
Device Catalogue Number699G
Device Lot NumberPDB848
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/14/2019
Initial Date Received by Manufacturer 09/30/2019
Supplement Date Received by Manufacturer10/29/2019
Initial Report FDA Received Date10/28/2019
Supplement Report FDA Received Date11/15/2019
Patient Sequence Number1
-
-