| Catalog Number |
04P54-25 |
| Medical Device Problem Code |
False Positive Result (1227)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
09/28/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.All available patient information was included.Additional patient details are not available.
|
| |
|
Event or Problem Description
|
|
The customer reported a false repeat reactive architect hbsag result on one patient.Results provided: (b)(6) 2019 sid (b)(6) = 1.38 / 1.29 / 1.43 s/co.Architect hbsag qualitative confirmatory testing = 102.364 (confirmed reactive).Patient was nonreactive on (b)(6) 2019 and (b)(6) 2019 (no data provided).No impact to patient management was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The evaluation of complaint data for the product and likely cause architect hbsag qualitative confirmatory lot 04330fn00 identified normal complaint activity and there are no trends for the product related to patient results.No customer returns were available for evaluation.Additionally, clinical testing of the architect hbsag qualitative confirmatory assay has been evaluated with a retained in-house kit of the same lot# 04330fn00 and met all specifications, confirming that this lot is meeting the product requirements.A review of the manufacturing documentation did not identify any issues associated with the customers observation.A review of the product labeling concluded that the issue is sufficiently addressed.Based on the investigation no product deficiency was identified for the architect hbsag qualitative confirmatory (lot# 04330fn00).
|
| |
|
Search Alerts/Recalls
|