Model Number TMICL13.2
Medical Device Problem Code Inadequacy of Device Shape and/or Size (1583)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 10/02/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Weight, ethnicity, race: unk.Health professional, occupation: unk.(b)(4).Lens work order search: no similar complaint type events reported for units within the same lot.(b)(4).
 
Event or Problem Description
The reporter indicated that the surgeon implanted a 13.2mm tmicl13.2 lens,-10.50/1.0/126 (sphere/cylinder/axis) into the patient's eye on (b)(6) 2019.On (b)(6) 2019 the lens was explanted due to excessive vault, and angle closure.No additional information has been provided.
 
Additional Manufacturer Narrative
Device evaluation : two lenses were returned in liquid, in a specimen cup.Visual inspection found a scratch on a haptic and residue on the lens surface.Refer to mdr 2019-01910 for fellow lens, claim#: (b)(4).
 
Additional Manufacturer Narrative
Event: the lens was implanted in the patient's right eye (od).Device evaluation should have been reported in supplemental mdr#1.Claim# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Common Device Name
PHAKIC TORIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
MDR Report Key9245636
Report Number2023826-2019-01909
Device Sequence Number11232450
Product Code QCB
UDI-Device Identifier00841542109093
UDI-Public00841542109093
Combination Product (Y/N)N
Initial Reporter StateNV
Initial Reporter CountryUS
PMA/510(K) Number
P030016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/30/2020
Device Model NumberTMICL13.2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/04/2019
Initial Date Received by Manufacturer 10/03/2019
Supplement Date Received by Manufacturer11/15/2019
12/31/2019
Initial Report FDA Received Date10/28/2019
Supplement Report FDA Received Date12/12/2019
12/31/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CARTRIDGE MODEL SFC-45, LOT UNK; FOAM TIP PLUNGER, LOT UNK; INJECTOR MODEL MSIPF, LOT UNK; CARTRIDGE MODEL SFC-45, LOT UNK; FOAM TIP PLUNGER, LOT UNK; INJECTOR MODEL MSIPF, LOT UNK
Outcome Attributed to Adverse Event Required Intervention;
Patient Age31 YR
-
-