MAUDE Adverse Event Report: STRYKER MEDTECH SONOPET 110V CONSOLE WITH FOOT PEDAL AND POLE; INSTRUMENT, ULTRASONIC SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER MEDTECH SONOPET 110V CONSOLE WITH FOOT PEDAL AND POLE; INSTRUMENT, ULTRASONIC SURGICAL Back to Search Results
Model Number 5450850000
Medical Device Problem Codes Inability to Irrigate (1337); Overheating of Device (1437)
Health Effect - Clinical Codes Burn(s) (1757); Burn, Thermal (2530)
Date of Event 09/30/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This record is a consolidation of records summarized as a part of the fda voluntary malfunction summary reporting program.Reported events: 1 event was reported for this quarter.Product return status: 1 device investigation type has not yet been determined.Additional information: 1 device is not labeled for single-use.1 device was not reprocessed or reused.
 
Event or Problem Description
This report summarizes 1 malfunction event in which the device reportedly experienced an irrigation issue that could lead to overheating and overheated.1 event had the patient receive a burn.
 
Additional Manufacturer Narrative
This record is a consolidation of records summarized as a part of the fda voluntary malfunction summary reporting program.Supplemental rationale: 1 previously reported event is included in this follow-up record.Product return status: 1 device was received.Event confirmation status: 1 reported event was confirmed.Evaluation results: 1 device was found to be affected by a fatigued irrigation/pinch valve.
 
Event or Problem Description
This report summarizes 1 malfunction event in which the device reportedly experienced an irrigation issue that could lead to overheating and overheated.1 event had the patient receive a burn.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SONOPET 110V CONSOLE WITH FOOT PEDAL AND POLE
Common Device Name
INSTRUMENT, ULTRASONIC SURGICAL
Manufacturer (Section D)
STRYKER MEDTECH
1131-1 higashi naganuma
inagi 206-0 802
JP  206-0802
MDR Report Key9245916
Report Number0008043928-2019-00017
Device Sequence Number430264
Product Code LFL
UDI-Device Identifier07613327291872
UDI-Public07613327291872
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)Y
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model Number5450850000
Device Catalogue Number5450850000
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/30/2019
Supplement Date Received by Manufacturer09/30/2019
Initial Report FDA Received Date10/28/2019
Supplement Report FDA Received Date07/29/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-