| Model Number |
5450850000 |
| Medical Device Problem Codes |
Inability to Irrigate (1337); Overheating of Device (1437)
|
| Health Effect - Clinical Codes |
Burn(s) (1757); Burn, Thermal (2530)
|
| Date of Event |
09/30/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This record is a consolidation of records summarized as a part of the fda voluntary malfunction summary reporting program.Reported events: 1 event was reported for this quarter.Product return status: 1 device investigation type has not yet been determined.Additional information: 1 device is not labeled for single-use.1 device was not reprocessed or reused.
|
| |
|
Event or Problem Description
|
|
This report summarizes 1 malfunction event in which the device reportedly experienced an irrigation issue that could lead to overheating and overheated.1 event had the patient receive a burn.
|
| |
|
Additional Manufacturer Narrative
|
|
This record is a consolidation of records summarized as a part of the fda voluntary malfunction summary reporting program.Supplemental rationale: 1 previously reported event is included in this follow-up record.Product return status: 1 device was received.Event confirmation status: 1 reported event was confirmed.Evaluation results: 1 device was found to be affected by a fatigued irrigation/pinch valve.
|
| |
|
Event or Problem Description
|
|
This report summarizes 1 malfunction event in which the device reportedly experienced an irrigation issue that could lead to overheating and overheated.1 event had the patient receive a burn.
|
| |
|