SMITH & NEPHEW, INC. LEGION TIB CONE ID 18 SHORT; PROSTH, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONST, CEMENTED, POLY/METAL/POLY Back to Search Results
Model Number 71935386
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 10/01/2019
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient was treated for pain and inflammation.It was discovered that the patient had an infection.All components were revised except for the patella.The patient had undergone a washout a year prior to this revision.
 
Additional Manufacturer Narrative
It was reported that the patient was treated for pain and inflammation.It was discovered that the patient had an infection.All components were revised except for the patella.The associated legion hinge knee baseplate, femoral wedge, pressfit stems, offset coupler, tibial wedge and tibial cone, used in treatment, were not returned for evaluation.Therefore a product analysis could not be performed.Our investigation including a review of the manufacturing records for the listed batches did not reveal any deviation from the standard manufacturing processes.A review of the complaint history for the listed parts revealed no prior complaints for the listed failure mode with the same batch number.The devices were sterilized according to sterilization release documentation from quality control.At this time, we have no reason to suspect that the products failed to meet any product specifications at the time of manufacture.A relationship, if any, between the devices and the reported incident could not be corroborated.No medical documents were received for investigation.Therefore no medical assessment can be performed at this time.Infection, a potential complication associated with any surgery, can occur and possible causes could include but not limited to contamination, patient reaction, and post-operative healing issue.Without the return of the actual product involved and no patient medical records available, our investigation of this report is inconclusive.Based on this investigation, the need for corrective action is not indicated.No further investigation is warranted for this complaint; however smith and nephew will continue to monitor for future complaints and investigate as necessary.Should the device or additional information be received, the complaint will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEGION TIB CONE ID 18 SHORT
Common Device Name
PROSTH, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONST, CEMENTED, POLY/METAL/POLY
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key9247986
Report Number1020279-2019-03764
Device Sequence Number2599660
Product Code JWH
UDI-Device Identifier00885556645383
UDI-Public00885556645383
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K162775
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number71935386
Device Catalogue Number71935386
Device Lot Number17BTX0002A
Initial Date Received by Manufacturer 10/01/2019
Supplement Date Received by Manufacturer03/11/2020
Initial Report FDA Received Date10/28/2019
Supplement Report FDA Received Date03/13/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-