| Catalog Number |
8002062022 |
| Medical Device Problem Code |
Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Codes |
Hyperemia (1904); Superficial (First Degree) Burn (2685)
|
| Date of Event |
10/01/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The burn allegedly appeared as a "mild sunburn" and no treatment was required.
|
| |
|
Event or Problem Description
|
|
It was alleged that while the patient was lying on the pad, they allegedly received a 1st degree burn from the temperature pump.
|
| |
|
Additional Manufacturer Narrative
|
|
After investigation, the alleged reddening of the patients skin was found to not likely be related to a defect with the product as no malfunction or product defect was alleged.A conversation with a clinical science liaison identified that this was likely reactive hyperemia related to compression force while laying on the pad.Device updated to temperature therapy pad.
|
| |
|
Event or Problem Description
|
|
It was alleged that while the patient was lying on the pad, they allegedly received a 1st degree burn.
|
| |
|