• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, COSTA RICA LTDA BRK¿ XS TRANSSEPTAL NEEDLE, 98 CM LENGTH; TROCAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, COSTA RICA LTDA BRK¿ XS TRANSSEPTAL NEEDLE, 98 CM LENGTH; TROCAR Back to Search Results
Model Number G407211
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Cardiac Perforation (2513)
Date of Event 10/08/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported pericardial effusion could not be conclusively determined.Per the ifu, cardiac perforation is a known risk during the use of this device.
 
Event or Problem Description
Related manufacturing reference: 3008452825-2019-00538.During an atrial fibrillation ablation procedure a pericardial effusion occurred.Following the completion of the left sided ablation the patient became hypotensive.A pericardial effusion was confirmed by echocardiography and a pericardiocentesis was performed to stabilize the patient.Once blood pressure was stabilized the patient was transferred to cardiac surgery to repair the perforation.The ablation procedure was not completed and there were no performance issues with any abbott device.The user stated the perforation occurred during transseptal puncture with the brk needle.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BRK¿ XS TRANSSEPTAL NEEDLE, 98 CM LENGTH
Common Device Name
TROCAR
Manufacturer (Section D)
ST. JUDE MEDICAL, COSTA RICA LTDA
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
MDR Report Key9249736
Report Number3008452825-2019-00537
Device Sequence Number13979291
Product Code DRC
UDI-Device Identifier05414734205153
UDI-Public05414734205153
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K122587
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/29/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/31/2020
Device Model NumberG407211
Device Lot Number6086887
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/08/2019
Initial Report FDA Received Date10/29/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
BRK¿ XS TRANSSEPTAL NEEDLE, 71 CM LENGTH
Outcome Attributed to Adverse Event Required Intervention;
Patient Age72 YR
Patient Weight92
-
-