| Model Number |
MI1000 MED-EL CONCERT |
| Medical Device Problem Code |
Output Problem (3005)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
06/23/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device has been explanted and should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Event or Problem Description
|
|
The recipient suddenly had no hearing perception with the left device.There was no reported event that caused the loss of benefit, including any trauma.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: damage to the active electrode, as might be caused by an external mechanical impact, was determined to be the root cause of device failure.The problems given in the recipient report appear to match the damage found.This is a final report.
|
| |
|
Event or Problem Description
|
|
The recipient suddenly had no hearing perception with the left device.There was no reported event that caused the loss of benefit, including any trauma.
|
| |
|