| Catalog Number |
06P23-25 |
| Medical Device Problem Code |
False Positive Result (1227)
|
| Health Effect - Clinical Code |
Fall (1848)
|
| Date of Event |
10/21/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Apoc incident (b)(4).Apoc labeling will be evaluated during the investigation as pertaining to the event.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2019, abbott point of care (apoc) was contacted by a customer regarding i-stat troponin cartridges that yielded a suspected discrepant troponin result of 0.29 on a (b)(6) year old female patient who fell prior to coming in to emergency department.There was no additional patient information at the time of this report.Return product is not available for investigation.(b)(6).At this time there is no reason to suspect a malfunction exists.The reporting decision was based on the limited information available that suggests the product was not performing within the variability of the assay.There is no evidence the patient suffered an mi.The investigation is underway.Per i-stat system manual: art: 714258-00q.Reportable range: 0.00 - 50.00.Reference range: 0.00 - 0.03 (represents the 0 to 97.5% range of results).Reference range: 0.00 - 0.08 (represents the 0 to 99% range of results).
|
| |
|
Additional Manufacturer Narrative
|
|
Apoc incident #: (b)(4).The investigation was completed on 12/04/2019.A review of the device history record (dhr) confirmed that the cartridge lot met finished goods (fg) release criteria.Retained cartridge testing met the acceptance outlined in appendix 1 of q04.01.003 rev.Ad (product complaint level 2 and level 3 investigation procedure).No deficiency has been identified for ctni cartridge lot g19253.
|
| |
|
Event or Problem Description
|
|
Na.
|
| |
|