| Model Number |
5873 |
| Medical Device Problem Codes |
Mechanical Problem (1384); Use of Device Problem (1670)
|
| Health Effect - Clinical Code |
Laceration(s) (1946)
|
| Date of Event |
10/04/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported while trying to position an unstable cervical spine, the mayfield osi coupler assembly adapter failed to tighten.Patient had an injury/laceration as a result of the malfunction.
|
| |
|
Event or Problem Description
|
|
It was reported while trying to position an unstable cervical spine, the mayfield osi coupler assembly adapter failed to tighten.Patient received a minor laceration.
|
| |
|
Additional Manufacturer Narrative
|
|
The equipment that was returned appeared to be in good condition.Testing showed that the device meets it specification; it locks in position and supports a load as expected.Unable to determine a reason why the users were unable to secure the device.No physical cause found.User error is suspected.
|
| |
|