| Model Number |
MMT-522NAP |
| Medical Device Problem Code |
Device Difficult to Program or Calibrate (1496)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/10/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the insulin pump received a button error.The customer¿s blood glucose level was unknown.The customer also reported that the time was moving forward.The customer was advised to discontinue the use of insulin pump and revert to backup plan.The device will be returned for analysis.
|
| |
|
Search Alerts/Recalls
|