This report summarizes malfunction events.A review of the events indicated that three (3) patient sample tests using the echo automated blood bank system produced unexpected negative results for specific antibodies.The results represent false negatives due to prior patient histories or, testing completed by other methods that demonstrated the presence of the specific antibodies.The malfunction did not show positive results as expected for the anti-e antibody on one (1) occasion when using the capture r ready id reagent.The malfunction did not show positive results as expected for the anti-c antibody on one (1) occasion when using the capture r ready id reagent.When the customer adjusted the well residual volume (wrv) the sample results were as expected.The malfunction did not show positive results as expected for the anti-jka antibody on one (1) occasion when using the capture r ready screen 3 reagent.Subsequent testing of reagent retention lots, reviews of design history records and reagent antigen validation testing of the initial bulk product used for commercial vialing showed acceptable results.Through post event tests and investigations, no specific causes were determined for the malfunctions.
|