COOK IRELAND LTD ZIMMON BILIARY STENT; FGE CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Model Number G21711
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation is still pending, a follow up mdr will be submitted to include the investigation conclusions.
 
Event or Problem Description
Event is fda mdr reportable based on the device malfunction reporting precedence for this device family for the issue of ¿stent kinked/bent'.Reported by the dm via phone call- while trying to put the stent onto the guidewire, the pig tail got kinked.The procedure was completed with a new device.
 
Event or Problem Description
Event is fda mdr reportable based on the device malfunction reporting precedence for this device family for the issue of ¿stent kinked/bent'.Reported by the dm via phone call while trying to put the stent onto the guidewire, the pig tail got kinked.The procedure was completed with a new device.
 
Additional Manufacturer Narrative
1 x zso-7-15 of lot number c1554918 was returned to cirl open in its original packaging.The device involved in the complaint was evaluated in the laboratory on 25th october 2019.A slight kink was noted at the 2nd and 5th tapered end porthole.No straightener was returned.A 0.035 wireguide passed through the stent fine.Prior to distribution all zso-7-15 are subjected to visual and functional checks to ensure device integrity.These inspections and functional checks are outlined in internal procedures in place in cirl.A review of the manufacturing records for zso-7-15 of lot number c1554918 did not reveal any discrepancies that could have contributed to this complaint issue.The review of relevant manufacturing records, confirms the failure mode has not previously occurred with the current lot number.Based on the information available to date, there is no evidence to suggest that there are any manufacturing issues associated with lot number c1554918.The instructions for use, ifu0045-6 which accompanies this device instructs the user to ¿visually inspect with particular attention to kinks, bends and breaks.If an abnormality is detected that would prohibit proper working condition, do not use.¿ also, the user is instructed by the instructions for use, ifu0045-7: ¿care must be exercised when straightening the pigtail curls* in order to avoid kinking or breaking the stent.¿ there is no evidence to suggest that the user did not follow the instructions for use.A definitive root cause cannot be determined, as the circumstances of use cannot be replicated in the laboratory and due to limited information.A possible cause of this complaint may be that the kink occurred while being straightened as it was being loaded onto the wire guide.Complaint is confirmed as the failure was verified in the laboratory.According to the information reported, the patient did not experience any adverse effects due to this occurrence.Complaints of this nature will continue to be monitored for potential emerging trends.
 
Additional Manufacturer Narrative
Investigation is still pending, a follow up mdr will be submitted to include the investigation conclusions.
 
Event or Problem Description
Event is fda mdr reportable based on the device malfunction reporting precedence for this device family for the issue of ¿stent kinked/bent'.Reported by the dm via phone call- while trying to put the stent onto the guidewire, the pig tail got kinked.The procedure was completed with a new device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZIMMON BILIARY STENT
Common Device Name
FGE CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
COOK IRELAND LTD
o halloran road
limerick
MDR Report Key9252171
Report Number3001845648-2019-00570
Device Sequence Number2104398
Product Code FGE
UDI-Device Identifier00827002217119
UDI-Public(01)00827002217119(17)211102(10)C1554918
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K851962
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/26/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Expiration Date11/02/2021
Device Model NumberG21711
Device Catalogue NumberZSO-7-15
Device Lot NumberC1554918
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/22/2019
Was the Report Sent to FDA? No
Event Location Hospital
Initial Date Received by Manufacturer 10/02/2019
Supplement Date Received by Manufacturer10/02/2019
10/02/2019
Initial Report FDA Received Date10/29/2019
Supplement Report FDA Received Date11/26/2019
12/24/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-