| Catalog Number |
1012448-15 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/07/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
G9: exemption number e2019001-permits numbering sequence to begin with 10000, to avoid duplication of report numbers due to process transition.There may be gaps in numbering for reports submitted during the transition period.The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents from this lot.The investigation determined the reported complaint appears to be related to operational context.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was performed to treat a moderately calcified, moderately tortuous left circumflex that was 90% stenosed.The lesion was pre-dilated with a traveler balloon then a xience sierra stent was deployed.For post-dilatation, a 2.75x15mm nc trek was used, however, the balloon ruptured during the first inflation at 10 atmospheres.A new non-abbott device was used to successfully complete the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.
|
| |
|