MAUDE Adverse Event Report: PENUMBRA, INC. POD PACKING COIL; HCG, KRD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. POD PACKING COIL; HCG, KRD Back to Search Results
Catalog Number RBYPODJ45
Medical Device Problem Code Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/26/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure to treat a type ii endoleak using a lantern delivery microcatheter (lantern) and pod packing coils (podjs).During the procedure, while attempting to advance a podj through the lantern with direct puncture, the lantern was kicked out due to the needle access being too close in proximity to the endograft.Subsequently, the physician decided to retract the podj.While retracting the podj through the lantern, the physician experienced resistance; therefore, the lantern containing the podj was removed.Upon removal, the lantern appeared kinked.The procedure was completed using a new lantern and podj.There was no report of an adverse effect to the patient.
 
Additional Manufacturer Narrative
Please note that the following section was inadvertently missed on the initial mfr report and is being included on this follow-up #01 mfr report: 3005168196-2019-02001.This report is associated with mfr report number: 3005168196-2019-02002.
 
Additional Manufacturer Narrative
Please note that the following section was incorrectly reported on the initial mfr report and is being corrected on this follow-up #02 mfr report:3005168196-2019-02001.1.Section d.Box 4.Expiration date this report is associated with mfr report number: 1.3005168196-2019-02002 h3 other text : placeholder.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POD PACKING COIL
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key9253914
Report Number3005168196-2019-02001
Device Sequence Number16756502
Product Code HCG
UDI-Device Identifier00814548016726
UDI-Public00814548016726
Combination Product (Y/N)Y
Initial Reporter CountryFR
PMA/510(K) Number
K141134
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/30/2019
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberRBYPODJ45
Device Lot NumberF90130
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 09/30/2019
Supplement Date Received by Manufacturer01/14/2005
01/14/2005
Initial Report FDA Received Date10/29/2019
Supplement Report FDA Received Date10/31/2019
12/31/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Date Report Sent to FDA01/01/2005
-
-