After the event it was found by the hospital that the flow sensor module was broken.As a disturbance of the expiratory flow measurement may affect all flow derived measurement values like the tidal volume, the minute volume, the flow curve and also the peep measurement as it depends on a flow trigger it can be assumed that a faulty flow sensor probably was root cause for the reported symptom that the peep value was not consistent with the set value.In this case the savina will post a high priority alarm "flow sensor inoperable" audible and visible.Additional alarms are posted if measurement values reach the alarm limits set by the user.It is recommended if not done already to replace the flow sensor.The exact proceeding how to exchange the flow sensor is described in the ifu.The number of similar cases, related to the same root cause, is within the expected range of the respective risk assessment and thus accepted.
|