Model Number PFSR120010
Medical Device Problem Code Calibration Problem (2890)
Health Effect - Clinical Code Injury (2348)
Date of Event 10/16/2019
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that after surgeon burred distal femur, he burred holes for femoral 4 in one cutting block.He then put the cutting block spikes into the burred holes and used the visualization tool to navigate cutting block.When the block was centered in the burr holes the cut was proud by about 10mm.The surgeon then navigated the block approx 10mm posterior from the burr holes and it indicated it was right on plan.So the burr holes were not in correct spot but approximately 10mm anterior.Noted that during initial burring of distal femur is was notice that the tracker arrays were very loose on the handpiece and would not let surgeon burr.Once tightened he was able to burr distal femur and continue.Surgery was completed using visualization tool to navigate to plan.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAVIO HANDPIECE TRACKER ARRAY
Common Device Name
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
MDR Report Key9254538
Report Number3010266064-2019-00124
Device Sequence Number2550353
Product Code OLO
UDI-Device Identifier00885556628577
UDI-Public00885556628577
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K152574
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberPFSR120010
Device Catalogue NumberPFSR120010
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 10/16/2019
Supplement Date Received by Manufacturer01/19/2021
Initial Report FDA Received Date10/30/2019
Supplement Report FDA Received Date01/20/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Weight136
-
-