• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC IRELAND RESOLUTE ONYX RX; STENT, CORONARY, DRUG-ELUTING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND RESOLUTE ONYX RX; STENT, CORONARY, DRUG-ELUTING Back to Search Results
Model Number RONYX35015UX
Medical Device Problem Codes Activation, Positioning or Separation Problem (2906); Device Dislodged or Dislocated (2923); Material Deformation (2976)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/12/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information: the device was inspected before use with no issues reported.Device evaluation summary: the stent had moved proximally on the balloon and was not positioned on the balloon between the marker bands as per specifications.Crimp impression were visible on the exposed balloon surface.The balloon folds were expanded.Deformation was evident to the entire stent with struts raised and stretched.Deformation was evident to the distal tip.Kinks were evident to both hypotube and distal shaft.No other damage was evident to the remainder of the device.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
During a procedure an attempt was made to use a resolute onyx rx coronary drug eluting stent.It was reported that the stent did not deploy, the stent got pushed back proximal to the balloon on the shaft of the delivery system.No patient injury was reported.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESOLUTE ONYX RX
Common Device Name
STENT, CORONARY, DRUG-ELUTING
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
MDR Report Key9254946
Report Number9612164-2019-04543
Device Sequence Number16614169
Product Code NIQ
UDI-Device Identifier00643169557079
UDI-Public00643169557079
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
P160043
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/19/2020
Device Model NumberRONYX35015UX
Device Catalogue NumberRONYX35015UX
Device Lot Number0009447967
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/04/2019
Initial Date Received by Manufacturer 10/18/2019
Supplement Date Received by Manufacturer11/25/2019
01/17/2020
Initial Report FDA Received Date10/30/2019
Supplement Report FDA Received Date12/09/2019
01/21/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-