TANDEM DIABETES CARE T:SLIM X2 INSULIN PUMP WITH BASAL-IQ TECHNOLOGY; AUTOMATED INSULIN DOSING, THRESHOLD SUSPEND Back to Search Results
Model Number 1000096
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Death (1802); Hypoglycemia (1912)
Event Date 10/03/2019
Event Type  Death  
Manufacturer Narrative
The device is expected to be returned; however, the device has not yet been received.A supplemental report will be submitted if the device is received.
 
Event Description
It was reported that the customer passed away.Cause of death is pending determination by coroner.Per coroner, customer's blood glucose was reportedly very low at around the time of death.No additional information was provided.
 
Manufacturer Narrative
On november 8, 2019, additional information was received from the medical examiner concluding that the patient¿s death was due to an accidental combination of xanax, oxycodone, methadone and gabapentin.Therefore, the cause of death has been determined to be prescription drug-related and unrelated to tandem¿s insulin pump or supplies.
 
Manufacturer Narrative
Tandem quality engineer reviewed pump data and concluded the following.There is no evidence that the pump experienced a malfunction or failure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
T:SLIM X2 INSULIN PUMP WITH BASAL-IQ TECHNOLOGY
Type of Device
AUTOMATED INSULIN DOSING, THRESHOLD SUSPEND
Manufacturer (Section D)
TANDEM DIABETES CARE
11075 roselle street
san diego CA 92121
Manufacturer Contact
mick trier
san diego, CA 92121
8584011451
MDR Report Key9256157
MDR Text Key164299127
Report Number3013756811-2019-75603
Device Sequence Number1
Product Code OZO
UDI-Device Identifier00853052007264
UDI-Public00853052007264
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P180008
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 10/30/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model Number1000096
Device Catalogue Number1000886
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/21/2019
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Initial Date Manufacturer Received 10/07/2019
Initial Date FDA Received10/30/2019
Supplement Dates Manufacturer Received11/08/2019
11/22/2019
Supplement Dates FDA Received11/08/2019
12/19/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/01/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Death;
Patient Age39 YR
-
-