MEDTRONIC, INC. PROGRAMMING HEAD; PROGRAMMER, PACEMAKER Back to Search Results
Model Number 2067L
Medical Device Problem Code Electrical Shorting (2926)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 07/17/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product analysis: analysis noted that the radiofrequency (rf) cable was pinched, which was causing a short.The shorted cable was causing the programmer to power down while attempting to start up.The rf head cable was replaced.The device passed all final functional tests.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the radiofrequency programmer head originally returned as an associated device subsequently tested out of specification during manufacturer¿s analysis.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING HEAD
Common Device Name
PROGRAMMER, PACEMAKER
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
MDR Report Key9257463
Report Number2182208-2019-01951
Device Sequence Number19302480
Product Code KRG
UDI-Device Identifier00721902290138
UDI-Public00721902290138
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P890003
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 10/30/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number2067L
Device Catalogue Number2067L
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/01/2019
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/22/2019
Initial Report FDA Received Date10/30/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-