| Model Number |
2067L |
| Medical Device Problem Code |
Electrical Shorting (2926)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
07/17/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product analysis: analysis noted that the radiofrequency (rf) cable was pinched, which was causing a short.The shorted cable was causing the programmer to power down while attempting to start up.The rf head cable was replaced.The device passed all final functional tests.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the radiofrequency programmer head originally returned as an associated device subsequently tested out of specification during manufacturer¿s analysis.There was no patient involvement.
|
| |
|