SMITH & NEPHEW, INC. LGN STYLUS 1MM; PROSTH, KNEE, PATE/FEMOROTIBIAL, SEMI-CONST, UNCEM, PORO, COAT, POLY/METAL/POLY
|
Back to Search Results |
|
Catalog Number 71434333 |
Device Problem
Material Separation (1562)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 10/10/2019 |
Event Type
malfunction
|
Event Description
|
It was reported that during surgery the screw came out during procedure.The part was removed from the patient.A smith & nephew back up was available.There was a 1-2 minute delay in procedure, but there was no harm to the patient.
|
|
Manufacturer Narrative
|
The associated device was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The arm bolt came out of the device, causing it to disassemble and become inoperable.The bolt and washer were not returned.The device was manufactured in 2014 and exhibits signs of extensive wear usage.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.
|
|
Search Alerts/Recalls
|
|
|