• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK IRELAND LTD ECHOTIP PROCORE HIGH DEFINITION ULTRASOUND BIOPSY NEEDLE; FCG KIT, NEEDLE, BIOPSY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK IRELAND LTD ECHOTIP PROCORE HIGH DEFINITION ULTRASOUND BIOPSY NEEDLE; FCG KIT, NEEDLE, BIOPSY Back to Search Results
Model Number G34785
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/02/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Pma/510(k) # k142688 the echo-hd-3-20-c device of lot number c1617971 involved in this complaint was returned for evaluation, with the original packaging.The packaging was open on receipt.With the information provided, a physical examination and document based investigation was conducted.The device related to this occurrence underwent a laboratory evaluation on the 16oct2019.There was a proximal break below the sheath extender.There was a unevenness felt approximately 8cm from the needle tip, the needle penetrated the target site 30 time per biopsy and this was potentially the cause of the needle breaking.As a result this possibly contributed to the unevenness felt in the needle during the lab evaluation there was a distal kink at the notch area, as per information provided, this was because the user accidentally bent the needle tip after the procedure.Prior to distribution, all echo-hd-3-20-c devices are subjected to functional checks and visual inspection to ensure device integrity.These inspections and functional checks are outlined in internal procedures in place at cirl.A review of the relevant manufacturing records (c1617971) revealed no discrepancies that could have contributed to this complaint.Upon review of complaints, this failure mode has not occurred previously with this lot number.Based on the information available to date, there is no evidence to suggest that there are any manufacturing issues associated with lot number c1617971.There is no evidence to suggest that the customer did not follow the instructions for use."visually inspect with particular attention to kinks, bends and breaks.If an abnormality is detected that would prohibit proper working condition, do not use." a definitive root cause could not be determined as the circumstances of use cannot be replicated in the laboratory.A possible root cause could be attributed to excessive force during the 60 scrolls that were made.Complaint is confirmed as the failure was verified in the laboratory.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.Complaints of this nature will continue to be monitored for potential emerging trends.
 
Event or Problem Description
The user advanced echo-hd-3-20-c/lot c1617971 into olympus's uct260.Biopsy of stomach smt was performed three times.On the first and the second biopsy, 60 scrolls (= 30 times of needle advancement per biopsy) were made.Then, the third biopsy was performed but the needle became unable to move by handle manipulation and the needle tip could not be retracted inside the sheath during the third biopsy.The procedure was continued with another device(boston's acquire) and completed successfully.There have been no adverse effects to the patient reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ECHOTIP PROCORE HIGH DEFINITION ULTRASOUND BIOPSY NEEDLE
Common Device Name
FCG KIT, NEEDLE, BIOPSY
Manufacturer (Section D)
COOK IRELAND LTD
o halloran road
limerick
MDR Report Key9260359
Report Number3001845648-2019-00572
Device Sequence Number17061757
Product Code FCG
UDI-Device Identifier00827002347854
UDI-Public(01)00827002347854(17)220529(10)C1617971
Combination Product (Y/N)N
Initial Reporter CountryJA
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/29/2022
Device Model NumberG34785
Device Catalogue NumberECHO-HD-3-20-C
Device Lot NumberC1617971
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/15/2019
Was the Report Sent to FDA? No
Distributor Facility Aware Date10/02/2019
Event Location Hospital
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/03/2019
Initial Report FDA Received Date10/31/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-