• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ELEKTA INC MOSAIQ; ACCELERATOR, LINEAR, MEDICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ELEKTA INC MOSAIQ; ACCELERATOR, LINEAR, MEDICAL Back to Search Results
Model Number MOSAIQ
Medical Device Problem Code Computer Software Problem (1112)
Health Effect - Clinical Codes No Known Impact Or Consequence To Patient (2692); No Information (3190)
Date of Event 04/23/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The manufacturer's investigation is on-going and further information will be provided once the investigation has completed.
 
Event or Problem Description
The customer reported that the gantry angle was different than the one defined in the setup field.Based upon the available information the clinician considers serious mistreatment to the patient.
 
Additional Manufacturer Narrative
The investigation was completed by conducting a thorough evaluation of the product and the reported information.Mosaiq worked as designed and intended.The user entry of the setup field data (e.G.Source angle) was not correct, and did not match the user assigned drr (digitally reconstructed radiograph) project which compromised image registration results.This issue is due to abnormal use.The customer is changing their procedure to eliminate initial data entry error by correctly sending set up fields from the treatment planning system (tps).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOSAIQ
Common Device Name
ACCELERATOR, LINEAR, MEDICAL
Manufacturer (Section D)
ELEKTA INC
100 mathilda place
5th floor
sunnyvale CA 94086
MDR Report Key9260498
Report Number2950347-2019-00019
Device Sequence Number19765908
Product Code IYE
UDI-Device Identifier00858164002220
UDI-Public00858164002220
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K183034
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,o
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberMOSAIQ
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 10/02/2019
Supplement Date Received by Manufacturer10/02/2019
Initial Report FDA Received Date10/31/2019
Supplement Report FDA Received Date01/27/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-