THORATEC SWITZERLAND GMBH CENTRIMAG MOTOR, OUS; PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE
|
Back to Search Results |
|
Model Number 201-10002 |
Device Problems
Vibration (1674); Infusion or Flow Problem (2964)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 09/07/2019 |
Event Type
Injury
|
Manufacturer Narrative
|
No further information was provided.A supplemental report will be submitted when the manufacturer's investigation is completed.
|
|
Event Description
|
It was reported that the centrimag motor was vibrating, a flow sensor fault occurred, the console display blanked then returned, and there was reduced flow from the system.The account exchanged to the backup system to which normal support resumed.No further information was provided.
|
|
Manufacturer Narrative
|
Section d3, g1, and g2: correction.No further information was provided.A supplemental report will be submitted when the manufacturer¿s investigation is completed.
|
|
Manufacturer Narrative
|
Manufacturer's investigation conclusion: the reported event of a flow sensor fault with a blank flow reading was confirmed via the log file and the investigation.A log file was extracted from the returned centrimag console (serial number (b)(6), evaluated under mfr # (b)(4).Per the log file, on (b)(6) 2019, the console was supporting the system at a speed of ~3800 rpm and ~2.85 lpm.An active s3 alarm was observed to have been caused by a ¿sf_ifd_shutdown_detected¿ sub-fault on (b)(6) 2019 at 9:14.A flow supply sub-fault was also observed within the same minute as a result of this event.After this event, speed dropped to ~3400 rpm and flow reading would have been blank with "--.--" on the console display, consistent with the reported information.The system was observed to be manually shut down at 11:19 on the same day and was not in use by the patient throughout the remainder of the log file.No other notable events were observed.The returned centrimag motor (serial number (b)(6) was tested on (b)(6) 2020.The motor was tested alongside the customer¿s returned console and flow probe, as well as a test flow probe.An s3 fault alarm occurred when either probe was in use during testing.Reports of similar events have been documented and corrective action (capa) has been initiated to investigate the issue.The motor was determined to be the root cause of the reported event of a blank flow reading with a flow sensor fault (correlating to an s3 alarm) and was sent to zurich for recalibration.Final disposition of the motors will be determined by the capa.Reports of similar events will continue to be tracked and monitored.No further information was provided.The manufacturer is closing the file on this event.
|
|
Search Alerts/Recalls
|
|
|