The patient was undergoing a microneurosurgery procedure using an artemis neuro evacuation device (artemis).During the procedure, the physician noticed that the artemis bident did not function when the device was activated.Additionally, it was noted that the artemis did not make any noise.The physician then decided to end the procedure and bring the patient back in the next day for an open procedure.There was no report of an adverse effect to the patient.
|
Results: the was no visible damage to the artemis.Conclusions: evaluation of the returned artemis revealed a functional device.The artemis was connected to a pump max.The pump max was powered, and the artemis was able to aspirate without an issue.In addition, the red button was depressed on the artemis and the bident wire was observed vibrating inside the tube.The root cause of the reported complaint could not be determined.Penumbra artemis devices are inspected at incoming quality control which includes a visual inspection as well as a verification of test results to ensure specifications for each output are met.The product lot number was not provided, therefore, the manufacturing records could not be reviewed.
|