BARD PERIPHERAL VASCULAR, INC. ULTRASCORE FOCUSED FORCE PTA BALLOON Back to Search Results
Model Number US35130630
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Vascular Dissection (3160)
Date of Event 09/12/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Manufacturing review: a manufacturing review was conducted and there was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the investigation is inconclusive, as no objective evidence has been provided to confirm any alleged deficiency with the device.Based upon the available information, the definitive root cause is unknown.Label review: the review of the ifu (instructions for use), indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.Medical device - expiration date: 05/2020.
 
Event or Problem Description
It was reported through the results of a clinical trial, that after an angioplasty procedure with a focused force percutaneous transluminal angioplasty balloon dilatation catheter, non-flow limiting dissection of target lesion was identified.It was further reported that surgical intervention was performed.
 
Additional Manufacturer Narrative
Additional information was received and dissection of the target lesion was removed from the system.Manufacturing review: a manufacturing review was conducted and there was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the investigation is inconclusive, as no objective evidence has been provided to confirm any alleged deficiency with the device.Based upon the available information, the definitive root cause is unknown.Label review: the review of the ifu (instructions for use), indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.( expiration date: 05/2020).
 
Event or Problem Description
It was reported through the results of a clinical trial, that after an angioplasty procedure with a focused force percutaneous transluminal angioplasty balloon dilatation catheter, non-flow limiting dissection of target lesion was identified.It was further reported that surgical intervention was performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRASCORE FOCUSED FORCE PTA BALLOON
Common Device Name
FOCUSED FORCE PTA BALLOON
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key9261674
Report Number2020394-2019-04342
Device Sequence Number2054622
Product Code PNO
UDI-Device Identifier00801741116926
UDI-Public(01)00801741116926
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K163420
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 11/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberUS35130630
Device Catalogue NumberUS35130630
Device Lot NumberGFCR3327
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 10/04/2019
Supplement Date Received by Manufacturer11/07/2019
Initial Report FDA Received Date10/31/2019
Supplement Report FDA Received Date11/11/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ASPIRIN, CLOPIDOGREL; ASPIRIN, CLOPIDOGREL; ASPIRIN, CLOPIDOGREL
Patient Age66 YR
-
-