Catalog Number 928858
Device Problem Device Markings/Labelling Problem (2911)
Patient Problem Hypoglycemia (1912)
Event Date 09/24/2019
Event Type  Injury  
Manufacturer Narrative
There were multiple lot numbers reported to be involved.The information for each lot number is as follows: medical device lot #: 9063755, medical device expiration date: 2024-03-31, device manufacture date: 2019-03-04.Medical device lot #: unknown, medical device expiration date: unknown, device manufacture date: unknown.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation and/or device history review, a supplemental report will be filed.
 
Event Description
Material no: 928858, batch no: 9063755, unknown.It was reported that during use of the syringe 0.3ml 31ga 8mm 10bag 500 pl/wg the scale markings at the top of the syringe are thicker and make it difficult to read and have the correct dose.Dex com device alarm went off with a glucose reading of less than 40.For the past 6 weeks mother of the child wakes up during the night to feed daughter sugar because of being consistently low.The following information was provided by the initial reporter: it was reported that the customer is finding a good amount with the scale markings thicker toward the top of syringe between 1/2unit marker and 1 1/2unit marker.Mom stated her child (daughter) using the syringe has been getting consistent low readings during the night for the past 6 weeks.09/24/2019 daughters dex com device alarm went off with a glucose reading of less than 40.For the past 6 weeks mom wakes up during the night to feed daughter sugar because of being consistently low.Mom feel the scales being thicker toward the top of the syringe is throwing the accuracy of the dose off.
 
Event Description
Material no: 928858 ; batch no: 9063755, unknown.It was reported that during use of the syringe 0.3ml 31ga 8mm 10bag 500 pl/wg the scale markings at the top of the syringe are thicker and make it difficult to read and have the correct dose.Dex com device alarm went off with a glucose reading of less than 40.For the past 6 weeks mother of the child wakes up during the night to feed daughter sugar because of being consistently low.The following information was provided by the initial reporter: it was reported that the customer is finding a good amount with the scale markings thicker toward the top of syringe between 1/2unit marker and 1 1/2unit marker.Mom stated her child (daughter) using the syringe has been getting consistent low readings during the night for the past 6 weeks.(b)(6) 2019 daughters dex com device alarm went off with a glucose reading of less than 40.For the past 6 weeks mom wakes up during the night to feed daughter sugar because of being consistently low.Mom feel the scales being thicker toward the top of the syringe is throwing the accuracy of the dose off.
 
Manufacturer Narrative
Investigation summary: no samples (including photos) were returned therefore the complaint could not be confirmed and the root cause is undetermined.Complaints received for this device and reported condition will continue to be tracked and trended.If samples are received in the future the complaint will be reopened for further investigation.A review of the device history record was completed for batch #9063755.All inspections and challenges were performed per the applicable operations qc specifications except as noted below.There was one (1) notification (b)(4) noted for damaged tips.As no samples and/or photo(s) were received the investigation concluded: unconfirmed: bd was not able to duplicate or confirm the customer¿s indicated failure as no samples or photos were returned.Based on the above, no additional investigation and no capa is required at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYRINGE 0.3ML 31GA 8MM 10BAG 500 PL/WG
Type of Device
PISTON SYRINGE
Manufacturer (Section D)
BD MEDICAL - DIABETES CARE
1329 west highway 6
holdrege NE 68949
MDR Report Key9261707
MDR Text Key167786652
Report Number1920898-2019-01224
Device Sequence Number1
Product Code FMF
UDI-Device Identifier00311917048154
UDI-Public00311917048154
Combination Product (y/n)N
PMA/PMN Number
K024112
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,other
Type of Report Initial,Followup
Report Date 12/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number928858
Device Lot NumberSEE H.10
Initial Date Manufacturer Received 09/25/2019
Initial Date FDA Received10/31/2019
Supplement Dates Manufacturer Received09/25/2019
Supplement Dates FDA Received12/05/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-