• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET 3I OSSEOTITE® IMPLANT 5 X 10MM; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET 3I OSSEOTITE® IMPLANT 5 X 10MM; DENTAL IMPLANT Back to Search Results
Model Number N/A
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Information (3190)
Date of Event 09/27/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
This report is being submitted to report (b)(4).The reported device has been received for evaluation.Investigation has not yet begun.Once investigation has been completed, a follow up medwatch will be submitted.
 
Event or Problem Description
It was reported that the implant at tooth location 46 was removed due to bone loss.
 
Event or Problem Description
No additional or corrected information to report.
 
Additional Manufacturer Narrative
This report is being submitted to relay additional information.The following sections are being reported: b4: date of this report was updated.D4: expiration date and unique identifier (udi) number were added.G4: date received by manufacturer was updated.G7: type of report was updated.H2: type of follow up was updated.H3: device evaluated by manufacturer was updated.H4: device manufacturer date was updated.H6: method code was updated to 3331 and 4109.H6: results code was updated to 213.H6: conclusions code was updated to 4310.H10: narrative/data was updated.The reported device was received for evaluation.The reported medical condition of bone loss was not confirmed.Visual inspection of the as returned product identified wear and debris about the implant threads.Functional testing to recreate the reported event could not be performed due to the nature of the device and event.A device history record (dhr) review was completed for the reported device lot.It was confirmed that all operations and inspections were executed as per applicable procedure.No deviations or non-conformance, which could have caused or contributed to the reported event was noted as part of the dhr.A complaint history review was completed for the reported device lot for similar event and no other complaint was identified.A definitive root cause for the reported event could not be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OSSEOTITE® IMPLANT 5 X 10MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
MDR Report Key9262618
Report Number0001038806-2019-01294
Device Sequence Number13923133
Product Code DZE
UDI-Device Identifier00844868013259
UDI-Public(01)00844868013259(17)190122(10)2014010304(241)OSS510
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K063286
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2019
Reporter Type Manufacturer
Report Source consumer,foreign,health profe
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 01/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/22/2019
Device Model NumberN/A
Device Catalogue NumberOSS510
Device Lot Number2014010304
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/04/2019
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 10/04/2019
Supplement Date Received by Manufacturer12/27/2019
Initial Report FDA Received Date10/31/2019
Supplement Report FDA Received Date01/06/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-