| Model Number |
134046 |
| Medical Device Problem Code |
Mechanics Altered (2984)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/16/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during an open procedure for enteritis, while clipping the vessel the surgeon able to squeeze the handle, however there were issue with loading and firing of the clips.The surgeon then used another device to resolve the issue in order to complete the case.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The instrument was received fully applied.Visual inspection of the instrument revealed no abnormalities.Functionally, the empty cartridge precludes firing evaluation; however, the interlock was engaged and properly prevented the jaw from approximating.In addition; each device is inspected during manufacturing to confirm the presence of a full clip complement.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.If information is provided in the future, a supplemental report will be issued.
|
| |
|